WEKO3
-
RootNode
アイテム
A new inlet system for microscale carbon and oxygen stable isotope analysis using dual inlet isotope ratio mass spectrometer at Niigata University, Japan
http://hdl.handle.net/10191/0002000226
http://hdl.handle.net/10191/00020002269763bd13-8a61-4f66-b4c2-23470543737f
名前 / ファイル | ライセンス | アクション |
---|---|---|
![]() |
|
Item type | 紀要論文 / Departmental Bulletin Paper(1) | |||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|
公開日 | 2022-01-21 | |||||||||||
タイトル | ||||||||||||
タイトル | A new inlet system for microscale carbon and oxygen stable isotope analysis using dual inlet isotope ratio mass spectrometer at Niigata University, Japan | |||||||||||
言語 | en | |||||||||||
言語 | ||||||||||||
言語 | eng | |||||||||||
キーワード | ||||||||||||
言語 | en | |||||||||||
主題Scheme | Other | |||||||||||
主題 | carbon and oxygen isotopes | |||||||||||
キーワード | ||||||||||||
言語 | en | |||||||||||
主題Scheme | Other | |||||||||||
主題 | MAT-253 inlet system | |||||||||||
キーワード | ||||||||||||
言語 | en | |||||||||||
主題Scheme | Other | |||||||||||
主題 | NIST standard measurements | |||||||||||
資源タイプ | ||||||||||||
資源 | http://purl.org/coar/resource_type/c_6501 | |||||||||||
タイプ | departmental bulletin paper | |||||||||||
著者 |
Satish-Kumar, Madhusoodhan
× Satish-Kumar, Madhusoodhan
× Kiran, Sasidharan
× Abe, Miyako
|
|||||||||||
抄録 | ||||||||||||
内容記述タイプ | Abstract | |||||||||||
内容記述 | A new Thermo Scientific^<TM> MAT-253 isotope ratio mass spectrometer was installed at Department of Geology, Niigata University, Japan under the MEXT Grant-in-Aid for Scientific Research on Innovative Areas. The mass spectrometer was set up for carbon and oxygen isotope measurement of small volume CO_2 samples. A new inlet vacuum line was constructed for transferring standard gas, to inlet externally prepared CO_2 gas samples and transfer CO_2 gas prepared in the online carbonate reaction system. The line is evacuated using a turbo molecular pump and pneumatically activated valves controlled by an electrical control board. The vacuum line also includes a reference CO_2 gas (Working Standard) reservoir and a laboratory standard CO_2 gas (Machine Standard) reservoir for routine analyses, as well as SF_6 reference gas reservoirs. The preparation line can hold high vacuum conditions without appreciable leaks for several hours, required for the completion of a single session of sample measurements. Long-term laboratory standard measurement results are reported here, which shows an average of δ^<13>C_<(V-PDB)> value of ‒29.796 ± 0.041‰ and δ^<18>O_<(V-SMOW)> value of 3.971 ± 0.062‰ (n = 44; 1σ). During a period of one year of analysis, a drift of 0.04‰ for both carbon and oxygen is observed for the reference CO_2 gas. NIST standards (NBS-20 limestone and NBS-21 graphite) and a laboratory diamond standard were analyzed to characterize the minimum volume of sample required to obtain carbon and oxygen isotopic composition with accuracy and precision <0.1‰. We also report here the routine analytical procedure for carbonate samples (calcite/aragonite and dolomite/siderite/magnesite) and elemental carbon (graphite and diamond). | |||||||||||
言語 | en | |||||||||||
書誌情報 |
en : Science reports of Niigata University. (Geology) 巻 36, p. 21-42, 発行日 2021-03 |
|||||||||||
出版者 | ||||||||||||
出版者 | Department of Geology, Faculty of Science, Niigata University | |||||||||||
言語 | en | |||||||||||
ISSN | ||||||||||||
収録物識別子タイプ | PISSN | |||||||||||
収録物識別子 | 1349-1237 | |||||||||||
書誌レコードID | ||||||||||||
収録物識別子タイプ | NCID | |||||||||||
収録物識別子 | AA11436808 | |||||||||||
出版タイプ | ||||||||||||
出版タイプ | VoR | |||||||||||
出版タイプResource | http://purl.org/coar/version/c_970fb48d4fbd8a85 |