Automatic Correction of Left-Ventricular Pressure Waveform Using the Natural Observation Method
http://hdl.handle.net/10191/20305
http://hdl.handle.net/10191/2030532838575-eb99-4302-b2ed-485fd22e5013
名前 / ファイル | ライセンス | アクション |
---|---|---|
![]() |
|
Item type | 学術雑誌論文 / Journal Article(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2012-10-31 | |||||
タイトル | ||||||
タイトル | Automatic Correction of Left-Ventricular Pressure Waveform Using the Natural Observation Method | |||||
タイトル | ||||||
タイトル | Automatic Correction of Left-Ventricular Pressure Waveform Using the Natural Observation Method | |||||
言語 | en | |||||
言語 | ||||||
言語 | eng | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | automatic correction | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | natural observation method | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | catheter manometer system | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | left-ventricular pressure waveform | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | lushing device | |||||
資源タイプ | ||||||
資源 | http://purl.org/coar/resource_type/c_6501 | |||||
タイプ | journal article | |||||
著者 |
Hori, Jun-ichi
× Hori, Jun-ichi× Saitoh, Yoshiaki× Kiryu, Tohru× Iijima, Taizo |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | The pressure waveforms indicated on a catheter manometer system are subject to serious distortion due to the resonance of the catheter itself, or the compliance of a particular transducer. Although several methods have been proposed for improving those characteristics, they have never been put into practice. We have focused on the transfer function of the catheter manometer, and made a pilot system, using the natural observation method. This method has been suggested as a means of studying the structure of the instantaneous waveform. In this manner, we were able to increase the bandwidth in the frequency domain and reduce the ringing in the time domain. Correction was performed automatically, using a step wave. Reproduction bf the waveform with a flushing device, was a task of equal simplicity, that allowed us to estimate the system parameters so that the response waveform became step-like. In the experiment, our system provided distortion-free left-ventricular pressure waveform measurements and exact evaluation of the cardiac pumping system. The values obtained came much closer to the original figures arrived at by the catheter-tip manometer system. | |||||
書誌情報 |
IEICE transactions on information and systems en : IEICE transactions on information and systems 巻 E75-D, 号 6, p. 909-915, 発行日 1992-11 |
|||||
出版者 | ||||||
出版者 | 電子情報通信学会 | |||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 09168532 | |||||
書誌レコードID | ||||||
収録物識別子タイプ | NCID | |||||
収録物識別子 | AA10826272 | |||||
権利 | ||||||
権利情報 | copyright@1992 IEICE | |||||
著者版フラグ | ||||||
値 | publisher | |||||
異版である | ||||||
関連タイプ | isVersionOf | |||||
識別子タイプ | URI | |||||
関連識別子 | http://www.ieice.org/jpn/trans_online/ |