WEKO3
アイテム
Improvement of the Time-Domain Response of a Thermodilution Sensor by the Natural Observation System (Special Section on Signal Processing and System Theory)
http://hdl.handle.net/10191/20222
http://hdl.handle.net/10191/20222e1dfe183-603e-4793-8194-10bf94634494
名前 / ファイル | ライセンス | アクション |
---|---|---|
![]() |
|
Item type | 学術雑誌論文 / Journal Article(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2012-10-24 | |||||
タイトル | ||||||
タイトル | Improvement of the Time-Domain Response of a Thermodilution Sensor by the Natural Observation System (Special Section on Signal Processing and System Theory) | |||||
タイトル | ||||||
言語 | en | |||||
タイトル | Improvement of the Time-Domain Response of a Thermodilution Sensor by the Natural Observation System (Special Section on Signal Processing and System Theory) | |||||
言語 | ||||||
言語 | eng | |||||
資源タイプ | ||||||
資源 | http://purl.org/coar/resource_type/c_6501 | |||||
タイプ | journal article | |||||
著者 |
Hori, Jun'ichi
× Hori, Jun'ichi× Saitoh, Yoshiaki× Kiryu, Tohru |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | When measuring the ejection fraction for the evaluation of the ventricular pumping function by means of the thermodilution technique, the slow response a conventional thermistor has caused it to be considered unsuitable, and fast thermistors have been proposed as an alternative. However, in this paper we propose improving the time-domain response of a conventional thermistor using a signal processing technique composed of a series of first-order high-pass filters which is known as the natural observation system. We considered the rise time of the thermistor in response to a step temperature change to effect correction for the measurement of the ejection fraction. The coefficients of the natural observation system were calculated by minimizing the square error between the step-response signal of the thermistor and the band-limited reference signal. In an experiment using a model ventricle, the thermodilution curve obtained from a conventional thermistor was improved using the proposed technique, thus enabling successful measurement of the ejection fraction of the ventricles. | |||||
書誌情報 |
IEICE transactions on fundamentals of electronics, communications and computer sciences en : IEICE transactions on fundamentals of electronics, communications and computer sciences 巻 E77-A, 号 5, p. 784-791, 発行日 1994-05 |
|||||
出版者 | ||||||
出版者 | 電子情報通信学会 | |||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 09168508 | |||||
書誌レコードID | ||||||
収録物識別子タイプ | NCID | |||||
収録物識別子 | AA10826239 | |||||
権利 | ||||||
権利情報 | copyright@1994 IEICE | |||||
権利 | ||||||
権利情報 | 本文データは学協会の許諾に基づきCiNiiから複製したものである | |||||
著者版フラグ | ||||||
値 | publisher | |||||
異版である | ||||||
関連タイプ | isVersionOf | |||||
識別子タイプ | URI | |||||
関連識別子 | http://www.ieice.org/jpn/trans_online/index.html | |||||
異版である | ||||||
関連タイプ | isVersionOf | |||||
識別子タイプ | URI | |||||
関連識別子 | http://ci.nii.ac.jp/naid/110003215808 |