2024-03-29T08:13:14Z
https://niigata-u.repo.nii.ac.jp/oai
oai:niigata-u.repo.nii.ac.jp:00002745
2022-12-15T03:35:57Z
423:424:425
453:454
植込み型センサを用いたハイパサーミア用温度測定システム
植込み型センサを用いたハイパサーミア用温度測定システム
A Temperature Measuring System with an Implantable Sensor for the Hyperthermia
斉藤, 義明
田中, 隆一
真柄, 睦
鈴木, 道斉
木竜, 徹
牧野, 秀夫
We have developed a temperature measuring system with an implantable sensor for the hyperthermia application. As the sensor does not require a battery, it can be implanted in a body for a long term and used repeatedly. The sensor consists of a coil and a quartz resonator whose resonance frequency varies with ambient temperature. The resonance frequency of the sensor can be measured sensitively from outside of the body by applying frequency modulated RF electromagnetic field. A microcomputer in the measuring system converts the resonance frequency to the corresponding temperature. Furthermore, the temperature data are displayed on a CRT display and stored in a floppy disk automatically. With this system, we have measured the temperature ranging from 35 to 50°C with errors less than ±0.1°C. The system operated properly while it was heated with 500W RF electromagnetic generator.
日本ME学会
1986-02
jpn
journal article
http://hdl.handle.net/10191/20184
https://niigata-u.repo.nii.ac.jp/records/2745
AN00016981
00213292
医用電子と生体工学
医用電子と生体工学
24
1
41
46
https://niigata-u.repo.nii.ac.jp/record/2745/files/24_1_41-46.pdf
application/pdf
2.7 MB
2019-07-29